Factory Default Values (Table 12-1)
S1-1
Closed
Excitation output is set to 5.0 V
S1-3
Open
High hystersis = 1.5 V
S1-4
Closed
Threshold is shifted to positive
S1-5
Open
No low pass filter
S1-6
Open
No pull-down resistor
S1-7
Open
243K pull-up to +excitation
S1-8
Open
Trigger edge = Negative (high to low transition)
S1-9
Open
No signal is internally connected to the input
S1-10
Open
Twice the line frequency/blank is not grounded
S2-C
Installed
Only "An LO" and "An HI" are locked-out
S3-A
Installed
Enables front-panel buttons
S4-A
Installed
Enables RECALL buttons
S5-A
Installed
Select for excitation output. You may program 5, 8.2 or 12.4 V

Configuration Defaults (Table 12-2)
FEATURE
DEFAULT DESCRIPTION
Operating Mode (Function)
Rate
Scale
Operating Mode (Function) Rate Scale Multiply by 1.00000
Offset
0.00000
Decimal Point
Fixed decimal point: FFFFFF
Setpoints
HI = 10000. LO = 10
Configuration
6 digit display. Each digit may be 0 (default) or 1. Configuration is detailed below.

6 Digits of Configuration
0
0
0
0
0
0
0
1

Enables analog output (4-20 mA)
Enables analog output (0-10 V)
0
1

Unlatched alarm
Latched alarm
0
1

Do not change low alarm
Change low alarm to high
0
1

Message handshake
Character handshake
0
1

9600 baud rate
1200 baud rate
0
1

Send alarm status character
Do not send alarm status character